Technical Specifications:
| Model |
AR-ELS20 |
|
Sample Unit |
Sample Racks |
Standard 192 sample positions (expandable, up to 552) |
| Pipetting |
10~1000μl |
| Dispensing System |
2 aspirating and dispensing probe (X-Y-Z movement) |
| Reagent Unit |
Reagent Position |
23 reagent racks, editable |
| Pipetting Range |
10~1000μl,  1μl stepping |
| Pipetting Precision |
(100μl)CV≤0.5% |
| Washing Unit |
Dispensing Time |
4 minutes to whole 96 to microplates |
| Washing Probes |
Two line 8-nozzle manifolds |
| Wash Containers |
Cleaning fluid(15L), waste water(25L), Buffer 1(5L), Buffer 2(2L),
Buffer 3(2L), Buffer 4(2L), with liquid level-sensing, liquid shortage and full alarm |
| Waste Containers |
25L with waste full sensor |
| Washing Residual |
<2μl |
| Reading Unit |
Reading Channels |
8 independent photometric channels, mono anddichromatic reading |
| Absorbance |
0.0~3.0OD |
| Spectral Range |
400~700nm |
| Optical Filters |
Multi‑filter reader (405–620 nm range, ≥6 filters) |
| Reading Accuracy |
±1% from 0~1.5OD, ±2% from 1.5~3.0OD |
| Incubation Unit |
Incubators |
4 independent units |
| Temperature Range |
RT to 45℃ |
| Temperature Accuracy |
±0.5℃ |
| Data Management |
Software System |
Above Windows 7 system |
| LIS System |
Bi-direction, support HL7 protocol |
| Results |
Absorbance and results reviewable by software |
| Printer |
External optional |
| Working Conditions |
Power Supply |
AC220V±10%, 50/60Hz; 110V±10%, 60Hz, 400W |
| Temperature |
10~30℃ |
| Humidity |
30%~80% |
| Size&Weight |
External Size(W*D*H) |
1235*705*875mm |
| Net Weight |
146kg |
| Package Size(W*D*H) |
1380*860*1415mm |
| Gross Weight |
220kg |
|