Specification
Testing Principles: The reflective photometer reads the color on the strip to calculate the concentration of each testing item and output a semi quantitative analysis result.
-Test items: GLU, BIL, SG, pH, KET, BLD, PRO, URO, NIT, LEU, VC, ACR
-Wavelength: 470nm, 525nm, 625nm
-Stray light: Relative intensity ≤ 1%
-Linearity: Linear correlation coefficient≥0.995
-Repeatability: Coefficient of variation CV ≤ 1%
-Stability: During 8 hours of operation, Coefficient of variation CV ≤ 1%
-Test throughput: 500 tests/hour
-Language: English
-Display: Large LCD touch screen, Resolution320×240
-Printer: Built-in thermal recorder, paper width 57mm
-Interface: Standard RS-232
-Storage: 2,000 results
-Light source: Super-high brightness LED white light source, with 30,000 hours lifespan
Accessories:
- Printer: Built-in or external laser printer
- Consumables: 100 test strips, reagents, calibrators, controls for startup
- UPS: Imported sine-wave UPS with 30 min backup
- Power: 220V, 50 Hz